News & press

Measuring breakthrough curves with ABR

The ABR is a dedicated breakthrough analyzer (BTA), fully automated and supplied with an integrated close-coupled mass spectrometer. The ABR is available on a range of configurations optimised to suit research-scale samples, with bed volumes from 2 cc to 20 cc. Up to 6 gas inlets are available as well as a dedicated purge stream. Flow rates are selected to suit the specific applications, and the ABR includes an ultra-low dead volume switching valve.

Options include an upgrade for operation at pressures to 50 bar, and an integrated vapor generator module for gas-vapor operation. The breakthrough data obtained is complementary to the adsorption-desorption isotherms measured with our IGA, XEMIS and IMI sorption analyzers.

ABR Key Features:

  • Fully automated breakthrough analyser
  • Optimised for research-scale sample sizes (bed sizes from 2 cc)
  • Ultralow dead volume automated switching (purge gas/process mixture)
  • Options for gas-gas, gas-vapor, vapor-vapor configurations
  • Optional high pressure (50 bar) configuration
  • Fully integrated close-coupled mass spectrometer

Leaks in the System

  • Any leak will cause a pressure change that is not due to sorption.
  • A pressure drop due to an outboard leak will appear as adsorption and a pressure increase due to an inboard leak will appear as desorption.
  • Routine vacuum hold and pressure hold tests should be conducted to ensure the system is leak tight.